CH1 广州清诚决策评估数字化报告

主评估窗口采用专家基准 1ms FFT。统计对象为 48 个现场 TXT 文件的 CH1 全量数据。比例均为频谱能量占比,不是时域电压占比。

一、总体指标

二、分型定义

三、分型统计

risk_classfile_countfile_ratiole_125k_medianle_125k_minle_125k_maxoutside_125k_medianle_200k_medianoutside_200k_medianabove_250k_medianabove_250k_minabove_250k_maxband_250_400k_medianband_400_1000k_medianmean_rms_medianmax_peak_median
A_low_frequency_stable3062.50%99.99%99.97%100.00%0.01%99.99%0.01%0.01%0.00%0.03%0.00%0.01%25.96673.243
B_transition_or_mixed714.58%82.26%81.53%95.11%17.74%83.95%16.05%14.91%4.12%15.55%3.06%11.85%18.04124.872
C_high_frequency_risk1122.92%48.56%28.97%70.96%51.44%53.32%46.68%43.36%24.47%59.92%8.92%34.44%17.75219.379

四、高频风险文件清单(优先现场复核)

filele_125koutside_125kle_200koutside_200kabove_250kband_250_400kband_400_1000kmean_rmsmax_peaktop_above_250kaction
1-机床移动 26-07-07 15-00-55.txt28.97%71.03%35.53%64.47%59.92%12.33%47.59%17.64518.76840.28%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 14-27-36.txt29.94%70.06%36.69%63.31%58.80%12.27%46.54%17.75219.37941.14%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 15-04-28.txt33.30%66.70%39.50%60.50%56.21%11.58%44.63%17.71119.07436.91%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 14-29-14.txt33.45%66.55%39.82%60.18%55.89%11.68%44.21%17.64418.76837.57%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 11-02-01.txt40.73%59.27%46.27%53.73%49.99%10.29%39.70%17.69119.07435.21%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 15-00-06.txt48.56%51.44%53.32%46.68%43.36%8.92%34.44%17.69519.07419.05%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-31-16.txt57.32%42.68%61.31%38.69%35.96%7.39%28.58%18.00419.6847.82%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-37-18.txt58.87%41.13%62.70%37.30%34.66%7.12%27.54%18.01822.4317.04%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-53-14.txt67.07%32.93%70.14%29.86%27.75%5.71%22.04%18.05920.2947.84%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-46-59.txt68.35%31.65%71.30%28.70%26.67%5.48%21.19%18.01219.6848.60%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-40-04.txt70.96%29.04%73.67%26.33%24.47%5.03%19.44%17.99419.3798.71%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险

五、风险文件高能窗口分布

filetop_event_counttop_start_min_stop_start_max_stop_above250_mediantop_above250_max
1-机床移动 26-07-07 15-00-55.txt201.1319.0400.4010.458
1-机床移动 26-07-07 14-27-36.txt200.4177.0470.4160.447
1-机床移动 26-07-07 15-04-28.txt200.83313.1690.3720.422
1-机床移动 26-07-07 14-29-14.txt200.0909.9370.3770.428
1-机床移动 26-07-07 11-02-01.txt200.46412.4110.3540.381
1-机床移动 26-07-07 15-00-06.txt200.1721.8320.1930.206
1-机床移动 26-07-07 16-31-16.txt205.1765.4760.0780.087
1-机床移动 26-07-07 16-37-18.txt204.6766.1990.0830.094
1-机床移动 26-07-07 16-53-14.txt202.4642.7300.0860.101
1-机床移动 26-07-07 16-46-59.txt202.1882.4360.0860.098
1-机床移动 26-07-07 16-40-04.txt201.7521.7770.0870.097

六、所有文件决策表

filerisk_classle_125koutside_125kle_200koutside_200kabove_250kband_250_400kband_400_1000kmean_rmsmax_peaktop_above_250kaction
1-机床移动 26-07-07 15-00-55.txtC_high_frequency_risk28.97%71.03%35.53%64.47%59.92%12.33%47.59%17.64518.76840.28%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 14-27-36.txtC_high_frequency_risk29.94%70.06%36.69%63.31%58.80%12.27%46.54%17.75219.37941.14%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 15-04-28.txtC_high_frequency_risk33.30%66.70%39.50%60.50%56.21%11.58%44.63%17.71119.07436.91%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 14-29-14.txtC_high_frequency_risk33.45%66.55%39.82%60.18%55.89%11.68%44.21%17.64418.76837.57%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 11-02-01.txtC_high_frequency_risk40.73%59.27%46.27%53.73%49.99%10.29%39.70%17.69119.07435.21%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 15-00-06.txtC_high_frequency_risk48.56%51.44%53.32%46.68%43.36%8.92%34.44%17.69519.07419.05%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-31-16.txtC_high_frequency_risk57.32%42.68%61.31%38.69%35.96%7.39%28.58%18.00419.6847.82%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-37-18.txtC_high_frequency_risk58.87%41.13%62.70%37.30%34.66%7.12%27.54%18.01822.4317.04%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-53-14.txtC_high_frequency_risk67.07%32.93%70.14%29.86%27.75%5.71%22.04%18.05920.2947.84%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-46-59.txtC_high_frequency_risk68.35%31.65%71.30%28.70%26.67%5.48%21.19%18.01219.6848.60%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-40-04.txtC_high_frequency_risk70.96%29.04%73.67%26.33%24.47%5.03%19.44%17.99419.3798.71%现场复核优先级P0:确认是否目标工况;500ksps有混叠/漏检风险
1-机床移动 26-07-07 16-54-37.txtB_transition_or_mixed81.53%18.47%83.27%16.73%15.55%3.19%12.35%18.04828.5340.79%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 16-48-34.txtB_transition_or_mixed81.85%18.15%83.56%16.44%15.29%3.14%12.15%18.04524.5671.21%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 16-59-18.txtB_transition_or_mixed82.13%17.87%83.84%16.16%15.01%3.09%11.93%17.98124.8720.91%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 16-58-14.txtB_transition_or_mixed82.26%17.74%83.95%16.05%14.91%3.06%11.85%18.04123.3460.81%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 16-45-10.txtB_transition_or_mixed83.52%16.48%85.06%14.94%13.88%2.85%11.03%17.97529.1440.38%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 16-51-45.txtB_transition_or_mixed84.70%15.30%86.25%13.75%12.78%2.63%10.15%18.06229.7550.31%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 15-16-02.txtB_transition_or_mixed95.11%4.89%95.57%4.43%4.12%0.85%3.28%17.96120.2943.31%现场复核优先级P1:可作为边界/过渡样本
1-机床移动 26-07-07 13-45-47.txtA_low_frequency_stable99.97%0.03%99.97%0.03%0.03%0.01%0.02%20.25545.6240.02%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 13-44-32.txtA_low_frequency_stable99.97%0.03%99.97%0.03%0.02%0.01%0.02%20.14345.0140.02%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 13-53-45.txtA_low_frequency_stable99.97%0.03%99.97%0.03%0.02%0.01%0.02%20.20143.1830.02%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 13-47-48.txtA_low_frequency_stable99.98%0.02%99.98%0.02%0.02%0.00%0.01%20.4111031.9670.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 13-48-46.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%20.4801517.8150.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 11-44-21.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%22.99955.6950.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 11-43-25.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%24.00959.3570.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-19-14.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%24.488662.3940.02%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-26-13.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%24.65369.1230.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-18-07.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%24.18359.9680.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-16-36.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%24.00060.5780.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-16-27.txtA_low_frequency_stable99.98%0.02%99.99%0.01%0.01%0.00%0.01%24.6311006.6380.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 16-57-30.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%25.53667.9020.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 16-58-03.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%25.49466.3760.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-11-19.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%26.27171.8700.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-02-39.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%26.57169.7330.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-02-00.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%26.49171.2590.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 16-59-12.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%25.89566.6820.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 16-59-56.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.01%26.03767.9020.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-03-28.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%26.686662.6990.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 11-09-45.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%27.32679.1940.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 10-59-46.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%27.87874.6160.01%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 11-34-54.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%28.49275.8370.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 11-42-11.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%29.43081.6360.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 10-56-34.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%30.26084.0770.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 10-47-58.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.01%0.00%0.00%29.88385.9080.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 10-52-27.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.00%0.00%0.00%31.19288.6550.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-04-03.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.00%0.00%0.00%26.7471715.2660.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-07 10-46-12.txtA_low_frequency_stable99.99%0.01%99.99%0.01%0.00%0.00%0.00%30.44383.7720.00%可作为500ksps低频可用性正样本
1-机床移动 26-07-06 17-29-07.txtA_low_frequency_stable100.00%0.00%100.00%0.00%0.00%0.00%0.00%197.266683.1470.00%可作为500ksps低频可用性正样本

七、结论

CH1 对 500ksps 的支持不是“全量无条件通过”,而是“多数文件强支持 + 少数高频风险文件需要现场解释”。今晚无法确认硬件和复采时,最关键的动作是明天优先复核 C 类文件是否属于目标碰/敲/切/挤。

输出文件:output/ch1_decision_report/